Manifold MC80 – STABILIZER is a second stage pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units. MC80 – STABILIZER is only second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds (such as MC80, MM90 or liquid oxygen tank (LOX). When there is a signal for pressure deviation in relation to the alarm settings, the alarm can easily be displayed on a gas alarm unit. It is also possible to send information to the central operation control. The stabilizer can be delivered with a pressure transmitter 4–20 mA or with a contact gauge.
The gas reduction unit MC80 – STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards.
MC80 STABILIZER INCLUDES THE FOLLOWING COMPONENTS:
MC80 Stabilizer Manifold
FOR A COMPLETE MC80 STABILIZER MANIFOLD ADD:
Alarm unit (included if ordered together with HP unit)
Ball valve DN15 with welding adaptors
03 | Product Information
Product Information
Item No.
Description
0727324
Denomination: MC80 STAB
Gas: O2
PRV: With pull-ring
Signalization: 4–20 mA
0727325
Denomination: MC80 STAB
Gas: Air
PRV: With pull-ring
Signalization: 4–20 mA
0727326
Denomination: MC80 STAB
Gas: N2O, CO2
PRV: With pull-ring
Signalization: 4–20 mA
0727339
Denomination: MC80 STAB
Gas: ALL
PRV: With pull-ring
Signalization: Contact gauge
0727340
Denomination: MC80 STAB
Gas: ALL
PRV: With pull-ring
Signalization: 4–20 mA
0727324
Denomination: MC80 STAB
Gas: O2
PRV: With pull-ring
Signalization: 4–20 mA
0727325
Denomination: MC80 STAB
Gas: Air
PRV: With pull-ring
Signalization: 4–20 mA
0727326
Denomination: MC80 STAB
Gas: N2O, CO2
PRV: With pull-ring
Signalization: 4–20 mA
0727339
Denomination: MC80 STAB
Gas: ALL
PRV: With pull-ring
Signalization: Contact gauge
0727340
Denomination: MC80 STAB
Gas: ALL
PRV: With pull-ring
Signalization: 4–20 mA
Regional information available in your language at
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow:
200 m3/h
Inlet maximal pressure:
20 bar (2 000 kPa)
Outlet nominal pressure:
4,5 bar (setting range 0,5–6 bar)
Inlet connection:
G3/4"F
Outlet connection:
G3/4"F
Pressure relieve valve:
6,8 bar
Pressure relieve valve pipe dimension:
ø 15 mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
We use cookies on our website to give you the most relevant experience by remembering your preferences and repeat visits. By clicking “Accept”, you consent to the use of ALL the cookies.
This website uses cookies to improve your experience while you navigate through the website. Out of these, the cookies that are categorized as necessary are stored on your browser as they are essential for the working of basic functionalities of the website. We also use third-party cookies that help us analyze and understand how you use this website. These cookies will be stored in your browser only with your consent. You also have the option to opt-out of these cookies. But opting out of some of these cookies may affect your browsing experience.