The pressure monitor makes sure that the lower distribution pressure for nitrous oxide compared to oxygen is kept. The lower nitrous oxide pressure will be maintained according to standards even when the emergency supply is used through quick connectors or central emergency supply. The pressure monitor is equipped with a digital pressure monitor unit monitoring the current gas pressures, and giving all the visual and acoustic alarms required by standards.
The signal to the gas alarm comes from pressure transmitters
The visual and audible signals can be sent to a manned area
The following gases are under surveillance: breathing Oxygen, Nitrous Oxide, Air and instrument Air
The alarm is indicated by an acoustic and and visual signal at the same time as the exact cause of the alarm is written on the display
This happens if the gas pressure rises above or sinks below the set maximum or minimum limits respectively
The pressure monitor is also equipped with bayonet coupling for breathing oxygen, nitrous oxide, breathing air, and instrument air.
O2, N2O, Air, Air–800, CO2, N2, VAC (all medical gases)
Number of gases
2 to 4 valves (DN15)
Working pressure
4–5 bar (breathing gases), 7–10 bar (instrumental gases), (-0,4)–(-0,9) bar (vacuum)
Maximum pressure
16 bar
Safety regulator capacity at 3 bar
150 l/min
Tube dimension
ø 15×1 mm, ø 22×1 mm
Emergency QC inlets
QC by national standards
Pressure gauges
0–16 bar
Pressure sensors
Transmitters 4–20 mA
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370
Gases
O2, N2O, Air, Air–800, CO2, N2, VAC (all medical gases)
Number of gases
2 to 4 valves (DN15)
Working pressure
4–5 bar (breathing gases), 7–10 bar (instrumental gases), (-0,4)–(-0,9) bar (vacuum)
Maximum pressure
16 bar
Safety regulator capacity at 3 bar
150 l/min
Tube dimension
ø 15×1 mm, ø 22×1 mm
Emergency QC inlets
QC by national standards
Pressure gauges
0–16 bar
Pressure sensors
Transmitters 4–20 mA
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370
06 | Accessories
Accessories
Item No.
Description
325197814
Gas: O2, Length: 1,3 m, Inlet conn.: SW NUT G3/8" – 6 mm, Outlet conn.: QC SS straight
325197815
Gas: N2O, Length: 1,3 m, Inlet conn.: SW NUT G3/8" – LH, Outlet conn.: QC SS straight
325197816
Gas: Air, Length: 1,3 m, Inlet conn.: SW NUT G3/8" – 8 mm, Outlet conn.: QC SS straight
325197817
Gas: Air–800, Length: 1,3 m, Inlet conn.: SW NUT G3/8", Outlet conn.: QC SS straight
325197814
Gas: O2, Length: 1,3 m, Inlet conn.: SW NUT G3/8" – 6 mm, Outlet conn.: QC SS straight
325197815
Gas: N2O, Length: 1,3 m, Inlet conn.: SW NUT G3/8" – LH, Outlet conn.: QC SS straight
325197816
Gas: Air, Length: 1,3 m, Inlet conn.: SW NUT G3/8" – 8 mm, Outlet conn.: QC SS straight
325197817
Gas: Air–800, Length: 1,3 m, Inlet conn.: SW NUT G3/8", Outlet conn.: QC SS straight
We use cookies on our website to give you the most relevant experience by remembering your preferences and repeat visits. By clicking “Accept”, you consent to the use of ALL the cookies.
This website uses cookies to improve your experience while you navigate through the website. Out of these, the cookies that are categorized as necessary are stored on your browser as they are essential for the working of basic functionalities of the website. We also use third-party cookies that help us analyze and understand how you use this website. These cookies will be stored in your browser only with your consent. You also have the option to opt-out of these cookies. But opting out of some of these cookies may affect your browsing experience.