Talk to the team

MC150 - STABILIZER

Capacity of 360 m3/h

Technical Overview

MC150 - STABILIZER is a second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds (such as MC80, MM90 or liquid oxygen tank (LOX). When there is a signal for pressure deviation in relation to the alarm settings, the alarm can easily be
displayed on a Gas alarm unit. It is also possible to send information to the central operation control.
The stabilizer can be delivered with either pressure transmitter 4-20 mA or with contact gauge
The gas reduction unit MC150 - STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards.
FOR A COMPLETE MC150 STABILIZER MANIFOLD ADD:
Alarm unit (included if ordered together with HP unit)

Product Information

Item No.
Description
325397706
Denomination: MC150 STAB, Air O2, RPV: Manual activation, Signalization Contact gauge
325397707
Denomination: MC150 STAB, Gas: O2, Air, N2, RPV: Manual activation, Signalization: 4-20 mA
325397706
Denomination: MC150 STAB, Air O2, RPV: Manual activation, Signalization Contact gauge
325397707
Denomination: MC150 STAB, Gas: O2, Air, N2, RPV: Manual activation, Signalization: 4-20 mA
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
360 m3/h
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
2× G1 1/2"F+soldering piece pipe ø 35 mm
Outlet connection
2× G1 1/2"F+soldering piece pipe ø 35 mm
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 35 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
360 m3/h
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
2× G1 1/2"F+soldering piece pipe ø 35 mm
Outlet connection
2× G1 1/2"F+soldering piece pipe ø 35 mm
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 35 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01

Technical Downloads

Find a distributor

linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram