Talk to the team

MC80 - DUPLEX

Capacity of 200 m3/h

Technical Overview

1. MC80 - HP UNIT
This module contains two regulators with pressure relieve valves and it is connected to two various cylinder banks with high pressure hoses. When the cylinder bank, which has been connected for operation, has been emptied the other duty side is automatically connected.

2. MC80 - STABILIZER
The stabilizer makes the operating pressure in the distribution system remain constant. The module contains two regulators with pressure relieve valves. Since the gas pressure is reduced in two steps the drop in pressure, when changing from the operating cylinder to the other bank of cylinders, is kept to a minimum. The unit is prepared for connection to a liquid oxygen supply tank (LOX).

3. GAS ALARM SYSTEM - based on the product variant
Alarm systems from GCE are user friendly solutions, with simple control and lots of extra functionality. It surveils electronically and the alarm sounds when the following happens:

1. Too high or too low distributing pressure,
2. Too high intermediate pressure,
3. Leakage on the reserve gas cylinder bank,
4. When change of operating side has been effected.

When connected to a liquid tank the following disturbances will be reported:

1. Too high or too low distribution pressure,
2. Too high intermediate pressure,
3. Leakage from the reserves,
4. When change of operating side has been effected.

The product is either equipped with an alarm system or a sensor only.
DUPLEX (MC80) INCLUDES THE FOLLOWING COMPONENTS:
MC80 – HP Unit
MC80 – Stabilizer
Sensors 4-20 mA. Alarm or just sensor based on item in column "Alarm"
Evacuating kits for collecting pipe
Shut-off valve for the distribution line
HP filters
FOR A COMPLETE DUPLEX MC80 STANDBY MANIFOLD ADD:
Collecting pipe set (high pressure valves and non-return valves, high pressure components)
High pressure hoses with safety wire
Gas name sign
Connection pipe 90 degree
Extension pipes if needed

Product Information

Item No.
Description
0732328
Denomination: DUPLEX 2×1, Gas: O2, RPV: Manual activation, Signalization: Alarm TMA9705
0732329
Denomination: DUPLEX 2×1, Gas: Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732330
Denomination: DUPLEX 2×1, Gas: N2O/CO2, RPV: Manual activation, Signalization: Alarm TMA9705
0732328
Denomination: DUPLEX 2×1, Gas: O2, RPV: Manual activation, Signalization: Alarm TMA9705
0732329
Denomination: DUPLEX 2×1, Gas: Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732330
Denomination: DUPLEX 2×1, Gas: N2O/CO2, RPV: Manual activation, Signalization: Alarm TMA9705
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
200 m3/h
TECHNICAL DATA - HIGH PRESSURE UNIT MC80
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
12 bar (setting range 10–16 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G3/4"F
Pressure relieve valve:
17 bar
Pressure relieve valve pipe dimension:
∅ 10 mm
Purge valves connection:
W21,8×1/14"M + pipe ∅ 15 mm
TECHNICAL DATA - STABILIZER MC80
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
G3/4"F
Outlet connection
G3/4"F
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
200 m3/h
TECHNICAL DATA - HIGH PRESSURE UNIT MC80
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
12 bar (setting range 10–16 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G3/4"F
Pressure relieve valve:
17 bar
Pressure relieve valve pipe dimension:
∅ 10 mm
Purge valves connection:
W21,8×1/14"M + pipe ∅ 15 mm
TECHNICAL DATA - STABILIZER MC80
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
G3/4"F
Outlet connection
G3/4"F
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01

Technical Downloads

Find a distributor

linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram