Talk to the team

Introduction

The humidifier for oxygen therapy is a device allowing an increase in the relative humidity in the oxygen supplied to the patient; this gas for medical use normally lacks a sufficient humidity grade to be physiologically tolerated.

The jar is made of polycarbonate, or polysulphone the cap with inlet adaptor of polypropylene and inlet connection of polypropylene + coated brass.

Gallery

Product Information

Item No.
Description
K294402
MEDIWET 134°C; 9/16" UNF F
K294432
MEDIWET 134°C; G 3/8" F
K293498
MEDIWET 134°C; M 12×1,25
K294435
MEDIWET 134°C; G 1/4"
K294402
MEDIWET 134°C; 9/16" UNF
K294416
MEDIWET 121°C; G 3/8"
K293491
MEDIWET 121°C; M 12×1,25
K294452
MEDIWET 121°C; G 1/4"
K294401
MEDIWET 121°C; 9/16" UNF
K294402
MEDIWET 134°C; 9/16" UNF F
K294432
MEDIWET 134°C; G 3/8" F
K293498
MEDIWET 134°C; M 12×1,25
K294435
MEDIWET 134°C; G 1/4"
K294402
MEDIWET 134°C; 9/16" UNF
K294416
MEDIWET 121°C; G 3/8"
K293491
MEDIWET 121°C; M 12×1,25
K294452
MEDIWET 121°C; G 1/4"
K294401
MEDIWET 121°C; 9/16" UNF
Regional information available in your language at
GCEGROUP.COM

Technical

Contents:
Only sterile water
Dimensions:
Height: 190 mm
Diameter: 57 mm
Width: 67 mm
Weight:
145 g
Capacity:
200 ml
Material:
Jar polycarbonate (autoclavable at 121 °C)
Jar polysulphone (autoclavable at 134 °C)
Lid and outlet hose nipple: polypropylene

Inlet nut: Chrome brass
Diffuser: Polyethylene
Hose: Silicone
Cleaning:
Water, non-abrasive detergent. Never use solvents.
Disinfection:
An alcoholic solution, or other solution compatible with the material according to the disinfectant manufacturer.
Autoclave:
Polycarbonate version 121°C for 20 minutes
Polysulphone version 134°C for 18 minutes
Special case:
Diffuser: Exchange at every cleaning - do not sterilise!
Maintenance:
Check that the humidifier is whole and airtight before use. Monthly exchange of gaskets is recommended. Exchange the diffuser when its micro perforations no longer exist.
Durability:
Minimum 20 autoclave cycles under the condition that all instructions accompanying the product are adhered.
Contents:
Only sterile water
Dimensions:
Height: 190 mm
Diameter: 57 mm
Width: 67 mm
Weight:
145 g
Capacity:
200 ml
Material:
Jar polycarbonate (autoclavable at 121 °C)
Jar polysulphone (autoclavable at 134 °C)
Lid and outlet hose nipple: polypropylene

Inlet nut: Chrome brass
Diffuser: Polyethylene
Hose: Silicone
Cleaning:
Water, non-abrasive detergent. Never use solvents.
Disinfection:
An alcoholic solution, or other solution compatible with the material according to the disinfectant manufacturer.
Autoclave:
Polycarbonate version 121°C for 20 minutes
Polysulphone version 134°C for 18 minutes
Special case:
Diffuser: Exchange at every cleaning - do not sterilise!
Maintenance:
Check that the humidifier is whole and airtight before use. Monthly exchange of gaskets is recommended. Exchange the diffuser when its micro perforations no longer exist.
Durability:
Minimum 20 autoclave cycles under the condition that all instructions accompanying the product are adhered.

Find a distributor