MM90 - HP UNIT SEMI AUTOMATIC

Capacity of 90 m3/h

Technical Overview

The MM90 HP unit medical manifold is intended for use in hospital pipeline system as medical gas source. Together with MM90, always use an alarm providing all alarms according to standard (like gas alarm). As 2nd stage is recommended to use a stabilizer.

MM90 includes the following components:
> MM90 HP unit Manifold
> Gas alarm
> Purge valves
> HP filters

FOR A COMPLETE MM90 HP UNIT MANIFOLD ADD:
> Collecting pipe set (high pressure valves, and non-return valves/high pressure components)
> Cylinder retaining brackets (included in gas cylinder collecting pipe set)
> High pressure hoses with safety wire
> Plug for close collecting pipeline
> Gas name sign
> Stabilizer
Gas Alarm is a standard accessory
The gas alarm gives visual and audible indication
The gas alarm is able to communicate with other equipment through relays
The manifold will deliver gas from the operating bank to the manifold pressure regulator until the cylinders are exhausted
At that point the supply will switch to the reserve bank and the exhausted bank can be replenished
The object gives uninterrupted gas supply

Product Information

Item No.
Description
0732314
Denomination: MM90 – HP unit 2×1, Gas: O2, RPV: Standard, Signalization: Alarm TMA6703
0732315
Denomination: MM90 – HP unit 2×1, Gas: Air, RPV: Standard, Signalization: Alarm TMA6703
0732316
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Standard, Signalization: Alarm TMA6703
0732320
Denomination: MM90 – HP unit 2×1, Gas: O2, Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732321
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Manual activation, Signalization: Alarm TMA9705
0727327*
Denomination: MM90 – HP unit 2×1, Gas: O2, N2O, Air, CO2, N2, RPV: Standard
*basic version without electric sensors
0732314
Denomination: MM90 – HP unit 2×1, Gas: O2, RPV: Standard, Signalization: Alarm TMA6703
0732315
Denomination: MM90 – HP unit 2×1, Gas: Air, RPV: Standard, Signalization: Alarm TMA6703
0732316
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Standard, Signalization: Alarm TMA6703
0732320
Denomination: MM90 – HP unit 2×1, Gas: O2, Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732321
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Manual activation, Signalization: Alarm TMA9705
0727327*
Denomination: MM90 – HP unit 2×1, Gas: O2, N2O, Air, CO2, N2, RPV: Standard
*basic version without electric sensors
Regional information available in your language at
GCEGROUP.COM

Technical

Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow:
90 m3/h
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
9 bar (setting range 9–15 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G1/2"M + soldering piece pipe ø 22 mm
Intermediate pressure relieve valve:
16 bar
Pressure relieve valve pipe dimension:
ø 10 mm
Purge valves connection:
W21,8×1/14"M + soldering piece pipe ø 10mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow:
90 m3/h
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
9 bar (setting range 9–15 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G1/2"M + soldering piece pipe ø 22 mm
Intermediate pressure relieve valve:
16 bar
Pressure relieve valve pipe dimension:
ø 10 mm
Purge valves connection:
W21,8×1/14"M + soldering piece pipe ø 10mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01